EN-27

Trooper Scale

GEOGRAPHICAL FACTOR (For Europe Only)

Press the G/N/T/

Menu button, gEO 19 is displayed. This is the current geographi-

cal adjustment value.

The geo factor includes settings from 0 to 31 and is used to compensate for
varying gravity at different geographical areas (complete geographical adjustment
settings are listed in the following table).

NOTE:

Only an authorized manufacturer’s representative or certified verification

personnel may make these changes.

Changing the geographical setting alters the calibration values.

Press the Print/

Units button until the desired geographical adjustment value is

reached. The factory default setting is gEO 19

SAVE
Press the G/N/T/

Menu button to end this block, SAVE is displayed.

Press the G/N/T/Menu button to save the geographical factor setting. The next
menu SEt uP is displayed.

or

Press the Print/

Units button to go back to the CAL menu without saving changes to

the geographical setting.

QUIT
To exit from the CAL menu, press the Print/

Units button to advance to Quit. Then

press the G/N/T/

Menu button to go back to the weighing mode.