FD Series EN-14

Span Calibration Procedure
• When CAL is displayed, press Yes to enter the calibration

sub-menu.

• When Span is displayed, press Yes to begin the Span

calibration process.

• Clear the pan if prompted to do so, then press Yes.....
• The display will show “--C--”, followed by the span

calibration weight value.

• If desired, press No to change to an alternate span

calibration weight value.

• Place the indicated calibration weight on the scale and

press Yes.

• The display will show “--C--”, then “done” and then

revert to weighing. Remove the calibration weight.

Note: The calibration procedure can be canceled at any
time by pressing the Exit button or turning the scale off.

Linearity Calibration Procedure
• When LIN is displayed, press Yes to begin the linearity

calibration process.

• Clear the pan if prompted to do so, then press Yes.
• The display will show “--C--”, followed by the first

calibration weight value.

• If desired, press No to change to an alternate first

calibration weight value.

• Place the indicated calibration weight on the scale

and press Yes.

• The display will show “--C--”, followed by the second

calibration weight value.

• Place the indicated calibration weight on the scale

and press Yes.

• The display will show “--C--”, then “done” and then

revert to weighing. Remove the calibration weight.

Note: The calibration procedure can be canceled at any
time by pressing the Exit button or turning the scale off.

End
• Press Yes to leave this sub-menu and go to the next.
• Press No to return to the top menu item in this sub-menu.