beautypg.com

Linearity calibration – Ohaus JR Series Electronic Balances User Manual

Page 15

background image

15

Linearity Calibration

Linearity Calibration

Linearity Calibration

Linearity Calibration

Linearity Calibration

1.

Turn Balance on. After zero reading,

press and hold

until CAL is

displayed, then release. Balance will

display SPAN.

2.

Press

and the display will show

LIN.

3.

Press

to start the Linearity Cali-

bration Procedure. When the

is released, C 0g will be displayed,

indicating that no weight should be in

the pan.

4.

Press

. The display will show

-C- followed by the value of the weight

which must be placed on the pan.

5.

Place the required weight on the pan.

ON