3 error codes, Remote interface reference – American Magnetics 185 & 186 Liquid Level Instruments (CE-Marked) User Manual
Page 60

50
Rev. 3
Remote Interface Reference
Error Codes
5.3 Error Codes
The Model 185/186 returns specific error codes for invalid commands and/
or arguments. If an error condition is returned, the command is not
processed and the configuration of the instrument is not modified. The
table below provides a list of error codes, their meaning, and any
associated limits.
Error Code
Meaning
Valid Range
−
1
LO setpoint out of range
0
≤
LO
≤
LENGTH
−
2
B setpoint out of range
0
≤
B
<
A
−
3
A setpoint out of range
B < A
≤
LENGTH
−
4
HI setpoint out of range
0
≤
HI
≤
LENGTH
−
5
Attempted to set or query for
LENGTH in PERCENT units mode
−
6
Invalid argument,
value
out of
maximum calibration range
1 cm
≤
value
≤
650 cm
−
7
INTERVAL setting out of range
0
≤
INTERVAL
≤
600 min
−
8
Unrecognized command
−
9
Invalid argument,
value
was negative
or non-numeric
−
0
Approximate calibration
factor
out of
range
0.1
≤
factor
≤
999.9
0RGHO
0RGHO
0RGHO