3 commands for querying the configuration, 4 command for returning a level measurement, Remote interface reference – American Magnetics 185 & 186 Liquid Level Instruments (CE-Marked) User Manual
Page 50
40
Rev. 3
Remote Interface Reference
Serial Command Set Reference
The SAVE command saves the HI, LO, A, B, INTERVAL, LENGTH, and
current remote units settings to permanent memory. Saved settings are
then recalled each time the power is turned off and then reapplied to the
instrument. If the configuration is changed from the front panel, the
settings are automatically saved to permanent memory.
5.1.4.3 Commands for querying the configuration
The HI, LO, A, B, INTERVAL, and LENGTH commands return the
current configuration of the instrument. Each return value is terminated
with
5.1.4.4 Command for returning a level measurement
The LEVEL command returns the liquid level in the current units selected
through the communication interface.
Command: HI
Function: Returns the HI
setpoint limit in the
current units
Returns:
Command: LO
Function: Returns the LO
setpoint limit in the
current units
Returns:
Command: A
Function: Returns the A
setpoint limit in the
current units
Returns:
Command: B
Function: Returns the B
setpoint limit in the
current units
Returns:
Command: INTERVAL
Function: Returns the fill
timer setting in
minutes
Returns:
Command: LENGTH
Function: Returns the active
sensor length in
the current units
Returns:
Command: LEVEL
Function: Returns the liquid
level in the current
units
Returns:
0RGHO
0RGHO
0RGHO