Calibration, Is the dielectric constant of a vacuum) – American Magnetics 185 & 186 Liquid Level Instruments (CE-Marked) User Manual
Page 28

18
Rev. 3
Calibration
Effects of dieletric shifts
To minimize the effects of shifts in the dielectric of the target liquid,
perform a closed dewar calibration (see page 21) at the expected operating
condition of the cryo-vessel. If this is not feasible, then calibrate the sensor
at atmospheric pressure and use the approximate calibration method to
compensate for the shift of the dielectric when the cryogenic liquid is
under pressure. For this type of approximate calibration, the reference
liquid
will be the target liquid at atmospheric pressure — see page 25 for a
detailed discussion of the approximate calibration method. If any
questions exist in regard to calibration issues, contact AMI for assistance
in determining the optimal calibration strategy.
Note
All references to “dielectric constant” herein refer to the unitless
relative dielectric to
ε
0
(
ε
0
is the dielectric constant of a vacuum).