Input:range – VXI VM4016 User Manual
Page 77
www.vxitech.com
VM4016 Command Dictionary
77
INPut:RANGe
Purpose
Sets the input range for one or more channels
Type
Setting
Command Syntax
INPut:RANGe
Command Parameters
*RST Value
100 for all channels
Query Syntax
INPut:RANGe?
Query Parameters
Query Response
Numeric ASCII value = 10 | 100
Description
The Input Range command selects the input range of one or more channels. The input
range may be set for ±10 V or ±100 V.
Note: The input offset is normalized to the ±10 V range. The actual input offset in the
100 V range is ten times the set value.
Examples
Command / Query
Response (Description)
INP:RANG 100,(@1,3,5,7)
(Sets the input range for Channels 1, 3, 5, and 7 to
100 V)
INP:RANG? 7
100 (Returns the set input range for Channel 7)
Related Commands