38
VM4016 Programming
B
RACKETING A
V
OLTAGE
In this example, an input voltage level will be bracketed for an over or under-voltage error
condition. The input voltage of 5.0 V will be monitored for an over-voltage of 5.25 V and an
under-voltage of 4.75 V. The error condition must be true for longer than 750 μs. An interrupt
will be generated if either of these conditions occur. Channel 1 will use for an over-voltage and
Channel 2 for an under-voltage. Channel 1 and Channel 2 positive (+) sides will be tied together
externally. Channel 1 and Channel 2 negative (-) sides will be grounded. The output interrupt will
not be latched but will be pulsed.
COMMANDS
DESCRIPTION
INP:RANG 10,(@1,2)
Selects ±10 V as the input range for Channel 1
and 2.
INP:DEB 75e-5
Sets the debounce time limit to 750 µs.
INP:MASK 1,(@1,2)
Enables Channel 1or 2 to generate an interrupt.
INP:MASK 0,(@3:16)
Disables Channels 3 through 16 from generating an
interrupt.
INP:POL NORM,(@1)
Selects Channel 1 to generate an interrupt when
Channel 1 is greater than the offset voltage.
INP:POL INV,(@2)
Selects Channel 2 to generate an interrupt when
Channel 2 is less than the offset voltage.
INP:OFFS +5.25,(@1)
Selects +5.25 Vas the offset (reference) voltage for
Channel 1.
INP:OFFS +4.75,(@2)
Selects +4.75 V as the offset (reference) voltage for
Channel 2.
OUTP:POL:EXT:INT NORM
Sets the external interrupt output to be active high.
Figure 3-2 and the explanation that follows illustrates what is occurring during this example.