INPUT:RANGE
The input range command selects the input range of one or more channels. The input range may
be either set for ±10 V or ±100 V. It is important to note that the input offset is normalized to
±10 V range. The actual input offset in the 100 V range is ten times the set value.
INPut:RANGe
Where
Where
channel list format supporting channels 1
through 16.
EXAMPLES
INPut:RANGe 100,(@1:16)
Sets the input range for Channels 1 through
16 to 100 V.
INPut:RANGe? 7
Returns the input range for Channel 7 as
100
100 V.
INP:RANG 10,(@4:6)
Sets the input range for Channels 4 through
6 to 10 V.
INP:RANG? 5
Returns the input range for Channel 5 as
10