Weighted bit values, The operation status group, Status:operation:condition – VXI VT1422A User Manual
Page 383: Comments

VT1422A Command Reference 381
STATus
Weighted Bit
Values
Register queries are returned using decimal weighted bit values. Enable registers can
be set using decimal, hex, octal, or binary. The following table can be used to help
set Enable registers using decimal and decode register queries.
Status System Decimal Weighted Bit Values
The Operation Status Group
The Operation Status Group indicates the current operating state of the VT1422A.
The bit assignments are:
STATus:OPERation:CONDition?
STATus:OPERation:CONDition?
returns the decimal weighted value of the bits set
in the Condition register.
Comments
•
The Condition register reflects the real-time state of the status signals.
The signals are not latched; therefore, past events are not retained in this register
(see STAT:OPER:EVENT?).
bit#
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
value
always 0
16,384
8,192
4,096
2,048 1,024
512
256
128
64 32 16 8
4
2
1
Bit #
dec value
hex value
Bit Name
Description
0
1
0001
16
Calibrating
Set by CAL:TARE and CAL:SETup. Cleared by
CAL:TARE? and CAL:SETup?. Set while *CAL?
executes and reset when *CAL? completes. Set by
CAL:CONFIG:VOLT or CAL:CONFIG:RES, cleared by
CAL:VAL:VOLT or CAL:VAL:RES.
1-3
Not used
4
16
0010
16
Measuring
Set when instrument INITiated. Cleared when
instrument returns to Trigger Idle State.
5-7
Not used
8
256
0100
16
Scan Complete
Set when each pass through a Scan List completed
(may not indicate all measurements have been taken
when TRIG:COUNT >1).
9
512
0200
16
SCP Trigger
An SCP has sourced a trigger event (future
VT1422A SCPs)
10
1024
0400
16
FIFO Half Full
The FIFO contains at least 32,768 readings
11
2048
0800
16
Algorithm Interrupted
The interrupt() function was called in an algorithm
12-15
Not used