Comments, Command sequence, Calibration:configure:voltage – VXI VT1422A User Manual
Page 266: Parameters

264 VT1422A Command Reference
CALibration
Comments
•
Related Commands: CAL:VAL:RES, CAL:STOR ADC
•
When Accepted: Not while INITiated
•
Send with VXIplug&play Function: hpe1422_cmd(...)
Command
Sequence
CAL:CONF:RES
connect reference resistor to Calibration
Bus
*OPC? or SYST:ERR?
must wait for CAL:CONF:RES to
complete
(now measure ref resistor with external DMM)
CAL:VAL:RES
Send measured value to module
CAL:STORE ADC
Store cal constants in non-volatile
memory (used only at end of complete
cal sequence)
CALibration:CONFigure:VOLTage
CALibration:CONFigure:VOLTage <range>,<zero_fs>
connects the on-board
voltage reference to the Calibration Bus. A measurement of the source voltage can
be made with an external multimeter connected to the H Cal and L Cal terminals on
the Terminal Module or the V H, V L,
Ω H, and Ω L terminals on the Cal Bus
connector when not using a terminal module. The
voltage level available when the
Parameters
Comments
•
The
±5% of one of the five following values:
0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, 16 V dc. This parameter may be
specified in millivolts (mv).
•
The expected FSCALE output voltage of the calibration source will be
approximately 90% of the nominal value for each range, except the 16 V range
where the output is 10 V.
•
When Accepted: Not while INITiated
•
Related Commands: CAL:VAL:VOLT, STOR ADC
•
Send with VXIplug&play Function: hpe1422_cmd(...)
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
range
numeric (float32)
see comments
volts
zero_fs
discrete (string)
ZERO | FSCale
none