292 VT1422A Command Reference
FETCh?
Use Sequence
MEM:VME:ADDR #H300000
MEM:VME:SIZE #H100000
1 megabyte (MB) or 262,144 readings
MEM:VME:STAT ON
×
°
(set up VT1422A for scanning)
×
TRIG:SOUR IMM
let unit trigger on INIT
INIT
program execution remains here until
VME memory is full or the VT1422A has
stopped taking readings
FORM REAL,64
affects only the return of data
FETCH?
Note
When using the MEM subsystem, the module must be triggered before executing the
INIT command (as shown above) unless an external trigger (EXT trigger) is used.
When using EXT trigger, the trigger can occur at any time.