Usage – VXI VT1422A User Manual
Page 380

378 VT1422A Command Reference
SOURce
•
To reduce the offset voltage at each dynamic strain "Buffered Output" channel:
1. Measure a Buffered Output channel with its bridge unstrained and place
the value in a variable arbitrarily called <offset_v>.
2. Send minus
For example: SOUR:VOLT <-offset_v>,(@10000)
•
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
SOUR:VOLT .041,(@10031)
correct a -41 mV offset at channel 31 of
VT1529A/B 00.
This manual is related to the following products: