Parameters, Comments, Usage – VXI VT1422A User Manual
Page 367: Sense:]strain:excitation

VT1422A Command Reference 365
[SENSe]
Parameters
Comments
•
The <ch_list> parameter must specify the channel used to sense the bridge
voltage, not the channel position on a Bridge Completion SCP.
•
Related Commands: SENSE:STRAIN:
…, SENSE:FUNC:STRAIN…,
MEAS:VOLT:EXCitation
•
*RST Condition: 3.9 V
•
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
STRAIN:EXC Meas_excV,(@107)
set measured excitation voltage for
channel 7
STRAIN:EXC Meas_excV,(@10022)
set excitation voltage for VT1529A/B
channel 0022
[SENSe:]STRain:EXCitation?
[SENSe:]STRain:EXCitation? (@<channel>)
returns the excitation voltage value
currently set for the sense channel specified by <channel>.
Parameters
Comments
•
The <channel> parameter must specify a single channel only.
•
Returned Value: Numeric value of excitation voltage. The data type is float32.
•
Related Commands: SENS:STRAIN:EXCitation, MEAS:VOLT:EXCitation
•
Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)
Usage
STRAIN:EXC? (@107)
query excitation voltage for channel 7
enter statement here
returns the excitation voltage set by
STR:EXC
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
excite_v
numeric (float32)
0.01 - 99
volts
ch_list
channel list (string)
100 - 15731
none
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
channel
channel list (string)
100 - 15731
none