beautypg.com

Sample, Subsystem syntax, Sample:timer – VXI VT1422A User Manual

Page 330: Parameters, Comments, Usage

background image

328 VT1422A Command Reference

Chapter 7

SAMPle

SAMPle

The SAMPle subsystem provides commands to set and query the interval between
channel measurements (pacing).

Subsystem Syntax

SAMPle

:TIMer <interval>
:TIMer?

SAMPle:TIMer

SAMPle:TIMer <interval>

sets the time interval between channel measurements.

It is used to provide additional channel settling time. See “Settling Characteristics”
on page 157

Parameters

Comments

The minimum is 40 µs. The resolution for is 2.5 µs.

If the Sample Timer interval multiplied by the number of channels in the

specified Scan List is longer than the Trigger Timer interval, at run time a
"Trigger too fast" error will be generated.

the SAMP:TIMER interval can change the effect of the

SENS:CHAN:SETTLING command. SENS:CHAN:SETT specifies the
number of times a channel measurement should be repeated for channels defined
in an algorithm. The total settling time per channel then is (SAMP:TIMER
<interval>) X (<chan_repeats> from SENS:CHAN:SETT)

When Accepted: Not while INITiated

Related Commands: SENSE:CHAN:SETTLING, SAMP:TIMER?

*RST Condition: Sample Timer for all Channel Lists set to 4.0E-5 seconds.

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

SAMPLE:TIMER 150E-6

Pace measurements at 150 µs intervals
channel to channel

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

interval

numeric (float32)

(string)

4.0E-5 to 16.3825E-3 |

MIN | MAX

seconds