Output:polarity, Parameters, Comments – VXI VT1422A User Manual
Page 318: Output:shunt, Usage, Output:polarity? output:shunt

316 VT1422A Command Reference
OUTPut
OUTPut:POLarity?
OUTPut:POLarity? (@<channel>)
returns the polarity on the digital output channel
in <channel>.
Parameters
Comments
•
The
•
Returned Value: returns one of NORM or INV. The type is string.
•
Send with VXIplug&play Function: hpe1422_cmdString_Q(...)
OUTPut:SHUNt
OUTPut:SHUNt <enable>,(@<ch_list>)
adds shunt resistance to one leg of bridge
on Strain Bridge Completion SCPs and the VT1529A/B Remote Strain Bridge unit.
This can be used for diagnostic purposes and characterization of bridge response.
Parameters
Comments
•
If
plug-on" error is generated.
•
Only one channel on any one VT1529A/B can be specified in
is because a single resistor is used to shunt each of a VT1529A/Bs 32 channels.
The
•
When Accepted: Not while INITiated
•
Related Commands: [SENSe:]FUNCtion:STRain
…, [SENSe:]STRain…
•
*RST Condition: OUTP:SHUNT 0 on all Strain SCP and RSC channels
•
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
OUTP:SHUNT 1,(@116:119)
add shunt resistance at channels 16
through 19
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
channel
string
100 - 163
none
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
enable
boolean (uint16)
0 | 1 | ON | OFF
none
ch_list
channel list (string)
100 - 15731
none