Usage, Output:current[:state, Parameters – VXI VT1422A User Manual
Page 316: Comments

314 VT1422A Command Reference
OUTPut
•
Returned Value: Numeric value of amplitude set. The data type is float32.
•
Related Commands: OUTP:CURR:AMPL
•
Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)
Usage
OUTP:CURR:AMPLITUDE? (@163)
Check SCP current set for channel 63
(returns +3.0E-5 or +4.88E-4)
OUTPut:CURRent[:STATe]
OUTPut:CURRent[:STATe] <enable>,(@<ch_list>)
enables or disables current
source on channels specified in <ch_list>.
Parameters
Comments
•
OUTP:CURR:STAT does not affect a channel's amplitude setting. A channel
that has been disabled, when re-enabled sources the same current set by the
previous OUTP:CURR:AMPL command.
•
OUTP:CURR:STAT is most commonly used to turn off excitation current to
four-wire resistance (and resistance temperature device) circuits during
execution of CAL:TARE for those channels.
•
When Accepted: Not while INITiated
•
Related Commands: OUTP:CURR:AMPL, CAL:TARE
•
*RST Condition: OUTP:CURR OFF (all channels)
•
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
OUTP:CURR OFF,(@100,108)
turn off current source channels 0 and 8
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
enable
boolean (uint16)
1 | 0 | ON | OFF
none
ch_list
channel list (string)
100 - 163
none