beautypg.com

Usage, Input:filter[:lpass]:frequency, Parameters – VXI VT1422A User Manual

Page 300: Comments

background image

298 VT1422A Command Reference

Chapter 7

INPut

When Accepted: Not while INITiated

Related Commands: INP:FILT:FREQ?, INP:FILT:STAT ON | OFF

*RST Condition: generally set to MIN. The VT1529A/B is set to 10 Hz.

Send with VXIplug&play Function: hpe1422_cmd(...)

Usage

INP:FILT:FREQ 100,(@100:119)

Set cutoff frequency of 100 Hz for first
20 channels

INPUT:FILTER:FREQ 2,(@15622)

Set cutoff frequency of 2 Hz for RSC
channel 15622

INPut:FILTer[:LPASs]:FREQuency?

INPut:FILTer[:LPASs]:FREQuency? (@<channel>)

returns the cutoff frequency

currently set for channel. Non-programmable SCP channels may be queried to
determine their fixed cutoff frequency. If the channel is not on an input SCP,
the query will return zero.

Parameters

Comments

The parameter must specify a single channel only.

This command is for programmable filter SCPs only.

Returned Value: Numeric value of Hz as set by the INP:FILT:FREQ command.

The data type is float32.

When Accepted: Not while INITiated

Related Commands: INP:FILT:LPAS:FREQ, INP:FILT:STATE

*RST Condition: generally set to MIN. The VT1529A/B is set to 10 Hz.

Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)

Usage

INPUT:FILTER:LPASS:FREQUENCY? (@155)

Check cutoff freq on channel 55

INP:FILT:FREQ? (@10024)

Check cutoff freq on RSC channel 0024

Parameter

Name

Parameter

Type

Range of

Values

Default

Units

channel

channel list (string)

100 - 15731

none