Diagnostic:test:remote:number, Parameters, Comments – VXI VT1422A User Manual
Page 289

VT1422A Command Reference 287
DIAGnostic
•
Returned Value: DIAG:REM:USER:DATA? returns an IEEE definite length
data block which represents an array of 894, int16 values.
•
RST Condition: Stored values not changed by *RST
•
Send with VXIplug&play Function: hpe1422_cmdInt16Arr_Q(...)
DIAGnostic:TEST:REMote:NUMber?
DIAGnostic:TEST:REMote:NUMber? <test_num>,<iterations>,(@<channel>)
executes a selected self-test number on a single Remote Signal Conditioning Unit
connected through the VT1539A SCP. See DIAG:TEST:REM:SELF? for details of
each test.
Parameters
Comments
•
The <test_num> parameter specifies the test to perform. For explanations of the
test numbers, See “DIAGnostic:TEST:REMote:SELFtest?” on page 288.
•
The <iterations> parameter specifies the number of times to perform a test.
•
The <channel> parameter may contain only any single channel number on a
Remote Signal Conditioning Unit. All channels on that RSCU will be tested.
•
Returned Value:
The data type for this returned value is int16.
•
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
test_num
numeric
1 - 5
none
iterations
numeric
1 - 32767
none
channel
channel list (string)
10000 - 15731
none
Value
Meaning
Further Action
0
test passed
None
1
number of times test
failed
Error information in FIFO buffer.
See error codes below.
-1
Couldn’t start
remote self-test
Query the Error Queue (SYST:ERR?)
See error messages starting on page 453.