Omega ILD44-UTP User Manual
Page 37
33
ALARM 2 ENABLE/DISABLE SUBMENU:
Press
d
5) Display flashes previous selection. Press
b until ENBL
displays to use Alarm 2.
Press
d
6) Display shows
STRD stored message momentarily and then
advances to
ABSo only if it was changed, otherwise press a to
advance to
ABSo Absolute/Deviation Submenu.
If
DSBL Alarm 2 Disabled was selected, all submenus of Alarm
2 will be skipped and meter advances to
LOOP Loop Break
Time Menu. If
ENBL Alarm 2 Enabled was selected, Output 2
will automatically Disabled, and reassigned as Alarm 2.
The remaining Alarm 2 menu items are identical to Alarm 1 Menu.
Modifying Alarm Settings will not reset the instrument.
3.2.9 Loop Break Time
Figure 3.8 Flow Chart for Loop Break Time
ENTER LOOP BREAK TIME MENU:
Press
a
1) Press
a, if necessary, until CNFG prompt appears.
Press
d
2) Display advances to
INPT Input Menu.
Press
a
3) Press
a, if necessary, until Display advances to LOOP Loop
Break Time Menu.
Press
d
4) Display advances to Loop Break Time
ENBL Enable or DSBL
Disable Submenu and flashes the previous selection.
LOOP BREAK ENABLE/DISABLE SUBMENU:
Press
b
5) Scroll through the available selections:
ENBL or DSBL.
Press
d
6) Display shows
STRD stored message momentarily and then
advances to
B.TIM Loop Break Time Value Submenu.
Loop Break is an additional safety feature intended to monitor the rate of change
of the process value, while approaching the SP1. It is strictly intended as an
additional warning system, therefore its use is entirely optional. An active Loop
Break will cause the Process Value digits to blink in a rotating pattern. If the
process value reaches the set point the blinking will stop and
B.TIM is completed
successfully, otherwise
BR.AL Break Alarm warning will flash, and Output 1 will
be turned off.