Omega ILD44-UTP User Manual
Page 26
22
ENTER READING CONFIGURATION MENU:
Press
a
1) Press
a, if necessary, until CNFG prompt appears.
Press
d
2) Display advances to
INPT Input Menu.
Press
a
3) Display advances to
RDG Reading Configuration Menu.
Press
d
4) Display advances to
DEC Decimal Point.
DECIMAL POINT SUBMENU:
Press
d
5) Display flashes previous selection for Decimal location.
Press
b
6) Scroll though the available selections and choose Decimal
location:
FFFF or FFF.F (also FF.FF and F.FFF — if PROC
Process type was selected in the Input Type Menu).
Press
d
7) Display shows
STRD stored message momentarily and then
advances to
TEMP Temperature Unit.
Decimal Point for Process Input Type is passive.
TEMPERATURE UNIT SUBMENU:
Press
d
8) Display flashes previous Temperature Unit selection.
Press
b
9) Scroll though the available selections to the Temperature Unit
of your choice:
°F or °C.
Press
d
10) Display shows
STRD stored message momentarily and then
advances to
FLTR Filter Constant.
FILTER CONSTANT SUBMENU:
Press
d
11) Display flashes previous selection for Filter Constant.
Press
b
12) Scroll though the available selections:
0001, 0002, 0004, 0008, 0016, 0032, 0064, 0128
Press
d
13) Display shows
STRD stored message momentarily only, if
change was made, otherwise press
a to advance to the next
menu.
If Process was selected in the Input Type Menu the display will
advance to
IN.RD Input/Reading Submenu, otherwise the
display advances to the
ALR1 Alarm 1 Menu.
The Filter Constant Submenu allows the user to specify the
number of readings stored in the Digital Averaging Filter.
For PID control select filter value 0001-0004. A filter value of 2 is
approximately equal to 1 second RC low pass time constant.