Omega Speaker Systems Dual Thermocouple Over the Internet iSE-TC User Manual
Page 56

Appendix E
iLog Error Messages
Error # Description
Note
-8003
User stopped logging readings.
-10005 Failed to find the iSE.
Ethernet cable is disconnected,
iSE is powered off, connections
across the firewall require longer
“connection to socket time out”
setting.
-10006 Windows socket was closed.
-10007 Windows socket error.
Wrong IP or wrong Port number
was used.
-10008 The iSE failed to respond to a request.
Wrong IP or wrong Port number
was used.
-10011 Response came empty.
No data was sent.
-10012 Device responded with
Possibly the iLog is configured for
"Serial Time Out" string.
wrong product model.
-10014 Terminal Server Mode when the Port is 1000. Try Port 2000 in iLog configuration.
-15100 Error on obtaining the temperature reading.
Possibly the iLog is configured for
wrong product model.