4.4 Telnet Setup

Set the Number of Connections to 1-5 other than 0, using telnet simulation program

connect to iSE. In continuous mode, the telnet teminal will receive continuous messages

from the iSE. In command mode, the command can be sent to query the iSE and get a

response back. Refer to Figure 3.3

Send remote reset: Telnet port 2002 will bring a terminal for admin password. After

typing the password and following with the end character, the message Admin. Login

Successful will be showed on the terminal. Now type "reset" following with return

character to reset the iSE. If no character is received within 20 seconds, the terminal will

be closed automatically.

Initiating the archive chart while Telnet is active will halt Telnet until the archive

chart is done.

4.5 HTTPget Program

The HTTPget software is used to send a single HTTP or TCP request to an iSE product.

In contrast, the telnet or Hyperterminal programs allow a continuous connection with

multiple requests to be sent to the iSE product.

Generally HTTPget is used for simply programming an IP address to the iSE or for

quickly obtaining a reading of from a device.

The iSE product must be configured from the configuration web page so that the "Server

Type" value is set to "Command" (This is positioned under the heading of Terminal

Server). Also the Number of Connections may need to be set to "0" to enable Port 1000

(Port 1000 is for access in a non-standard terminal mode).

To use Port 2000 access (where "2000" is the value stored in "Port"), the Number of

Connections should be set to "2" for general usage. The value of 2 can later be

changed to a value from 1 to 5 depending on needs for secure access or fault tolerance.

Whenever Terminal Server service (using Port 2000 by default) is required, the Number

of Connections must be set to a value from 1 to 5. The Terminal Server mode is the

recommended mode for the most reliable connection when operating with NEWPORT

software or with other programs supporting TCPIP communications.

The Port 1000 access can be used with NEWPORT software and may be needed with

some iSE products when you need to view readings from the web page while

simultaneously collecting data through TCPIP communications.