beautypg.com

Omega Speaker Systems Dual Thermocouple Over the Internet iSE-TC User Manual

Page 50

background image

PART 7

APPROVALS INFORMATION

7.1 CE APPROVAL

This product conforms to the EMC directive 89/336/EEC amended by 93/68/EEC,

and with the European Low Voltage Directive 72/23/EEC.

Electrical Safety EN61010-1:2001

Safety requirements for electrical equipment for measurement, control and laboratory.

Basic Insulation

Pollution Degree 2

Dielectric withstand Test per 1 min

Input Power to Ethernet Output:

1500Vac

Input Power to Relays:

1500Vac

Ethernet Output to Relays:

1500Vac

Measurement Category I

Category I are measurements performed on circuits not directly connected to the Mains

Supply (power). Unit measures Temperature.

Transients Overvoltage Surge (1.2/50uS Pulse)

Input Power:

500V Transients Overvoltage

Ethernet:

1500V Transients Overvoltage

Note: The ac/dc power adaptor must have Safety Qualified Agency Approvals

for CE with Double Insulation rating.

The ac/dc power adaptor is 9Vdc.

The minimum output current rating is 500mA.

EMC EN61000-6-1:2001 (Immunity) and EN61000-6-3:2001 (Emmissions)

Immunity requirements for residential, commercial and light-industrial environments

EMC Emissions

Table 1, Class B

EMC Immunity

Table 1: Enclosure

Table 2: Signal Lines Ports

Table 3: Dc input/Dc output Ports

EMC EN61326:1997 + and A1:1998 + A2:2001

Immunity and Emissions requirements for electrical equipment for measurement, control

and laboratory.

EMC Emissions

Table 4, Class B of EN61326

EMC Immunity

Table 1 of EN61326

Note:

I/O lines / sensor cables require shielded cables and these cables must be located

on conductive cable trays or in conduits.

Refer to the EMC and Safety installation considerations (Guidelines) of this manual for

additional information.

7.2

FCC

This device complies with Part 15, Subpart B, Class B of the FCC rules.

46