2 setting offsets (offst) – Omega RD260A User Manual
Page 62

6 - 11
6. SET MODE
6.2
Setting Offsets (OFFST)
When recording and showing data using the current input as the reference, measured
values can be corrected within 10% of the span. For example, assume the reading is
1.998 V for the current level of input. If you want to change the reading to 2.000 V, set
0.002 V as the offset. This menu opens only when offset settings protection is turned
on in the SET UP mode.
Key operation
Indication
Discription
Enter the SET mode by pressing the [DISP] key for three
seconds. Using the [
] key, show [
] and then
press the [ENT] key.
Using the [
] key, select the channel to be configured and
then press the [ENT] key. Configurable channels are those
whose measuring range is VOLT, TC, RTD or SCL.
Type the value of offset. Use the [
] key to move through
the digits and the [
] key to select a numeral. When setting
is complete, press the [ENT] key. Note that set the offset
within 10% of the recording span or scale.
[
] is displayed and the setting is then completed.
Press the [ENT] key to set another channel; or press the
[DISP] key for three seconds to return to the OPERATION
mode.
NOTE
If you press the [ESC] key in the middle of setting, the indication returns to the
[
] and already set data will be aborted.
6.2 Setting Offsets (OFFST)
DISP
ESC
3 sec
CH UP
CH UP
FEED
DISP
ESC
3 sec
CH UP
CH UP