Part 3 - keypad configuration – Omega FD6003 User Manual
Page 26

3-10
second). Valid ranges for this entry are 0.00 to 30.00 
FPS. Two useful equations that relate liquid velocity to 
volume: 
GPM = 2.45 X I.D.
2
X FPS
FPS = ( GPM X 0.408) / I.D.
2
I.D. in inches
n If a UNITS code for metric measurement was made
the I.D. value will be entered in MPS (meters per 
second). Valid ranges for this entry are 0 to 10.00 
MPS. Two equations that relate liquid velocity to 
volume are 
LPM = 0.047 X I.D.
2
X MPS
MPS = ( LPM X 21.28) / I.D.
2
I.D. in inches
Note: Attempting to set a FULL SCALE value of less than 
0.5 FPS [0.15 MPS] may result in an unstable transmitted 
output. If flows are typically lower than this range, the 
LOW FLOW FILTERS and higher DAMPing values may 
be required. 
UNITS
Utilized to set engineering units of measure. There are 
twelve different selections possible. The Table 3.1 lists 
the entry code number, flow rate unit of measure and 
totalizer unit of measure. 
Note: After changing the UNITS value, it may be 
necessary to change other configuration values 
accordingly. For example, FULL SCALE, I.D., HIGH/LOW 
ALARM are influenced by the UNITS entry. 
Note: Flow already accumulated will not be correctly 
compensated for if the UNITS of measure changes. 
PART 3 - KEYPAD CONFIGURATION