Part 2 - transducer installation – Omega FD6003 User Manual
Page 14

2-4
1. Pipe Preparation:
For permanent silicone adhesive mounting, after 
determining the transducer location, some attention 
must be given to the pipe condition. Before the 
transducer head is bonded to the pipe surface, an 
area slightly larger than the flat surface to the 
transducer head (black rectangle) must be cleaned to 
bare metal. This means the removal of all paint rust, 
and scale. Some minor pipe pitting will not cause 
problems, as the acoustic couplant will take up the 
voids. In the case where plastic pipe is used, remove 
all paint and grease so that a smooth, dry surface is 
exposed. 
2. Transducer Mounting:
The transducer center line is designed to mount 
parallel to the pipe center line. The groove in the 
transducer body will allow the 1/2” stainless steel 
strap that was enclosed with the meter to align the 
transducer properly on the pipe. DO NOT mount the 
transducer on bends, elbows or fittings. Every effort 
should be made to mount the transducer parallel to 
the axis of the pipe as well as flat on the pipe. The 
transducer cable should run in the “down-stream” 
direction of liquid flow. See Figure 2.2.
In horizontal pipe runs, mount the transducer 
between 2 and 4 o’clock from the top—12 o’clock 
position; prepare the pipe surface as described. 
Finish the surface with some emery paper and then 
wipe the surface with trichlorenthylene to thoroughly 
degrease the contact surface in a area slightly larger 
than the flat surface of the transducer. 
For permanent mounting, use a good silicone based
PART 2 - TRANSDUCER INSTALLATION
Intrinsic Safety 
Installations 
Installations requiring intrinsic safety should refer to 
the Appendix drawings covering these applications.
