Part 2 - transducer installation – Omega FD6003 User Manual
Page 13

2-3
permanently mounted. Under flowing liquid
conditions, adequate signal is indicated when the 
Signal Strength indicates between 000125 and 
000200 counts. 
For proper operation, there cannot be air voids 
between the traducer face and pipe. The space must 
be filled with a material which is a good transmitter of 
sound energy such as: 
SILICONE GREASE: Dow Corning 111 R or 
comparable (-100 to +450 F.) The material must be 
suitable not to flow at temperature of pipe. Used for 
temporary survey installations and portable flow 
meters. 
SILICONE RUBBER: Dow Corning 732-RTV R.
Excellent for permanent bonding. This adhesive is a 
recommended bonding agent and easily removable. 
INSTALLATION AND PIPE PREPARATION
The cable from the transducer is provided with
either dual-coaxial cables, flexible nylon conduit or 
PVC coated steel conduit with a 1/2” NPT fitting. The 
coaxial cable was ordered from the factory at a 
specific length UNDER NO CIRCUMSTANCES 
should the coaxial cable be lengthened as this 
may de-tune the circuitry and influence 
performance.
Installation of the small-pipe transducers follow the
same procedures as the standard pipe type. The only
difference is that the small pipe transducers utilize an
integral pipe clamping mechanism with two
opposing sensing heads and the standard pipe units 
use a stainless steel strap.
PART 2 - TRANSDUCER INSTALLATION
Acoustic Couplant 
Types 
Small Pipe 
Transducers 