beautypg.com

Accuracy information transfer characteristics, Voltage output – National Instruments NI 6014 User Manual

Page 74

background image

Appendix A

Specifications

© National Instruments Corporation

A-5

NI 6013/6014 User Manual

Accuracy Information

Transfer Characteristics

Relative accuracy (INL)......................... ±3 LSB, typ

DNL ....................................................... ±2 LSB, typ

Monotonicity .......................................... 15 bits

Offset error

After calibration .............................. ±372

µV max

Before calibration ........................... ±250 mV max

Gain error (relative to internal reference)

After calibration .............................. ±75 ppm

Before calibration ........................... ±22,700 ppm

Voltage Output

Range ..................................................... ±10 V

Output coupling...................................... DC

Output impedance .................................. 0.1

Ω max

Current drive .......................................... ±5 mA max

Protection ............................................... Short-circuit to ground

Power-on state (steady state).................. ±250 mV

Initial power-up glitch

Magnitude ....................................... ±6.0 V

Duration .......................................... 4 ms

Nominal Range (V)

Absolute Accuracy

% of Reading

Offset

(µV)

Temp Drift

(%/°C)

Absolute

Accuracy at

Full Scale

(µV)

Positive FS

Negative FS

24 Hours

90 Days

1 Year

10

–10

0.0154

0.0174

0.0196

1,873

0.0005

3,835

This manual is related to the following products: