C and Low-Level DAQ Functions

When using a C-based environment, several steps are needed to configure
the SCXI-1125 for multiplexed scanning. The following procedure outlines
the steps for programming with the low-level DAQ function calls:

1.

Prepare the SCXI-1125 settings by either loading the original SCXI
configuration settings using

SCXI_Load_Config

, or by specifying

the gain and filter settings using

SCXI_Set_Gain

and

SCXI_Configure_Filter

.

2.

Use

SCXI_SCAN_Setup

to specify the module scan list, the start

channel of each module, and the number of channels to scan on each
module.

SCXI_SCAN_Setup

accepts an array of start channels and an

array of the number of channels to scan in each module. It is not
possible to repeat channels or use nonsequential channels using

SCXI_SCAN_Setup

.

3.

Next, use

SCXI_MuxCtr_Setup

to program the E Series DAQ device

with the correct number of channels multiplexed per scan. This
number must match the total number of channels programmed in
step 2.

You are now ready to acquire the channel data with the E Series DAQ
device. If you are using a multifunction E Series DAQ device, you can use

SCAN_OP

to perform the scanning operation. After scanning, convert the

binary data to voltage data using

SCXI_Scale

. Refer to the NI-DAQ User

Manual for additional information on scanning with DAQ devices.

Using Software for Parallel Scanning

Performing scanning operations in parallel mode requires no special
channel strings or function calls for setting up channel sequencing as is
required in multiplexed mode. Scanning the SCXI-1125 channels on a
differentially configured DAQ device is done as if there were no
SCXI-1125 module connected. The only requirement is that you must
configure the module for parallel mode in MAX as described in Chapter 1,

About the SCXI-1125

.

LabVIEW and Parallel Mode

In LabVIEW, the SCXI-1125 configuration settings are automatically
passed from MAX. LabVIEW can also set the SCXI-1125 configuration
parameter, operating mode, to parallel or multiplexed programmatically
by using the Set SCXI Information VI. In Measurement Studio, set the
operating mode using MAX as described in Chapter 1,

About the

SCXI-1125

.