Troubleshooting
84
If you cannot resolve a problem, call LifeScan Healthcare Professional Line:
1 800 524-7226 (USA)
1 888 353-0800 (Canada)
Error 6
System Error
The bedside unit
detected a problem
with the electronics.
1.Wipe contact points in lens
area with cloth moistened
with water to remove
residual bleach. Dry
thoroughly, but
NOTE: Before cleaning
contact points, compare
color of test strip confirma-
tion dot to Color Chart on
test strip bottle. If glucose
level is very high or very low,
take appropriate steps to
monitor patient and provide
medical care.
• Repeat test with new
test strip.
• Make sure test strip lot
number (and code)
entered at bedside unit
matches number on test
strip bottle.
2.If you continue to get any
error message, obtain a
laboratory result. Take
appropriate steps to
monitor patient and
provide medical care while
awaiting laboratory result.
3.Contact LifeScan for help.