Lifescan ONE TOUCH BASIC BLOOD GLUCOSE MONITORING SYSTEM User Manual
Page 54

MESSAGE
PROBLEM
WHAT TO DO
4) There was not enough
Repeat the test with a new
blood on the Test Strip.
Test Strip and a larger drop
of blood.
5) The Meter was used in
Move the Meter out of
very bright light.
direct light and repeat the
test with a new Test Strip.
6) The Check Strip procedure
Repeat the Check Strip test.
was incorrect.
7) The Meter may not be
Call LifeScan Technical
operating correctly.
Services.
47
LSI-97-050 OTB Eg 7/25/97 11:09 PM Page 47