Figure 5.4 on, Replacing the elbow pads - model ih6077a only – Invacare Deluxe Three-Position IH6065 series User Manual
Page 24
SECTION 5—MAINTENANCE
24
FIGURE 5.4 Replacing the Recline Mechanism Guard
Replacing the Elbow Pads - Model IH6077A Only
NOTE: For this procedure, refer to FIGURE 5.5.
1. Remove two hex head screws that
secure the existing elbow pad to the
elbow pad bracket.
2. Remove existing elbow pad and
discard.
3. Using the existing hex head screws,
secure the new elbow pad to the elbow
pad bracket.
FIGURE 5.5 Replacing the Elbow Pads -
Adjusting Gas Spring Recline Cylinder Assembly -
Model IH6077A Only
NOTE: For this procedure, refer to FIGURE 5.6 on page 25.
1. Remove the mounting screw which secures the cable to the back assembly.
2. Remove the retaining ring from the clevis pin.
3. Remove mounting bolt and locknut securing the gas spring recline cylinder to the
back frame.
4. Recline the back until cylinder can be removed from mounting bracket easily
(approximately a 45° angle).
5. Remove gas spring recline cylinder.
6. Loosen the locknut on the bottom of the cylinder (Detail ʺAʺ).
7. Hold the bottom bracket and turn the cylinder until cable becomes taut.
NOTE: If the cylinder was not locking when the lever was actuated, turn cylinder clockwise. If the
cylinder was not locking when the lever was released, turn cylinder counterclockwise.
DO NOT Remove
This Bolt
Bolt A
Bolt B
Bolt C
Bolt D
Recline Mechanism
Guard
Footrest
Side Frame
Seat Cushion
Rear of Chair
Side View
Front of Chair
Hex Head Screws
Elbow Pad
Bracket
Elbow
Pad