beautypg.com

Graco TEXSPRAY 1030 User Manual

Page 9

background image

309662

9

Troubleshooting (Continued)

PROBLEM

CAUSE

SOLUTION

Displacement pump output

low on downstroke or both

Siphon hose strainer is clogged

Clean strainer

low on downstroke or both

strokes

Piston packings worn or damaged

Replace packings. See page 12.

Material leaks into wet-cup

Intake valve ball (216) not seating

properly

Clean and service intake valve. See

page 12.

Clutch (9) worn or damaged

Replace clutch. See page 18.

Air leak at pump foot valve or

siphon tube elbow

Be sure foot valve is tight each day.

Be sure seal in elbow is in place and is

not worn or damaged.

Material leaks into wet-cup

Loose wet-cup (202)

Tighten wet-cup enough to stop leakage

Throat packings worn or damaged

Replace packings. See page 12.

Displacement rod (201) worn or

damaged

Replace displacement rod. See page 12.

Intermittent material flow out

of gun

Material supply is low or empty

Refill and prime pump. See Startup proce-

dure in operator manual. Check material

supply often to prevent running pump dry.

Air control on gun is set too high.

Reduce the air flow through the gun by

turning the air control knob on the back of

the gun.

Hose too long or material too

coarse

Change to a shorter hose or less coarse

material.

Heavy or poorly atomized mate-

rial when gun is first triggered
(lack of soft start)

Fluid pressure was not allowed to

fully self relieve after releasing gun

trigger

Wait longer between releasing gun trigger

and re--triggering.

(

)

Gun was not fully triggered immedi-

ately when squeezed, causing gun

air passages to clog with material.

Always squeeze trigger fully to prevent

just air from turning on, which starts pump

and pressurizes hoses. Disassemble and

clean gun.

Air valve failure

Check and replace as needed

Air flow sensor dirty or damaged

Check air filter.
Clean air flow sensor (see page 4) or re-

place as needed.

Pump does not shut off after

gun trigger is released

Air leaks in the system after the air

flow sensor

Check and tighten all connections be-

tween the air flow sensor and the gun.

Air leaks in gun

Check gun for excessive air leaks and re-

pair.
See manual 309661

Air flow sensor dirty or damaged

Check air filter.
Clean air flow sensor (see page 4) or re-

place as needed.

Premature prime valve fail-

ures

Pump does not shut off after gun

trigger is released.

Read “Pump does not shut off after gun

trigger is released” and fix appropriate

problem.
Repair or replace Prime Valve.
See page 19.