2640a ac voltage measurement specifications – Fluke NetDAQ 2645A User Manual
Page 219

Specifications
2640A Specifications
A
A-9
Table A-11. 2640A DC Voltage Accuracy Specifications
Accuracy, 3
σ
±
(% input + V)
18
°
C to 28
°
C
-10
°
C to 60
°
C
Range
90 Day
1 Year
1 Year
Slow
Fast
Slow
Fast
Slow
Fast
90 mV
.01%+7 µV
.01%+17 µV
.013%+8 µV
.013%+18 µV
.042%+18.2µV .042%+44.2µV
300 mV
.01%+15 µV
.01%+30 µV
.013%+17 µV
.013%+35 µV
.042%+39 µV
.042%+78 µV
3V
.01%+.1 mV
.01%+.2 mV
.013%+.15 mV
.013%+.2 mV
.042%+.26 mV .042%+.52mV
30V
.01%+1.5 mV
.02%+3 mV
.013%+1.7 mV
.026%+3.5 mV
.042%+3.9 mV .084%+7.8mV
150/300V
.01%+15 mV
.04%+30 mV
.013%+17 mV
.052%+35 mV
.042%+39 mV
.168%+78 mV
Note 300V range applies to channels 1 and 11 only.
2640A AC Voltage Measurement Specifications
A-14.
Tables A-12 to A-14 provide 2640A specifications for the ac voltage measurement
function.
Table A-12. 2640A AC Voltage General Specifications
Specification
Characteristic
Input Impedance
1 M
Ω
in parallel with 100 pF
Maximum Crest Factor
3.0 Maximum
2.0 for rated accuracy
Crest Factor Error
For non-sinusoidal input signals with crest factors
between 2 and 3 and pulse widths
≥
100 µs, add
0.2% to the accuracy specifications.
Common Mode Rejection
80 dB minimum at dc, 50 Hz/60 Hz
±
0.1%, 1 k
Ω
imbalance, Slow Rate