Table A-14. 2640A AC Voltage Accuracy Specifications

1 Year Accuracy + (%input + V)*

Range

Frequency

18

º

C to 28

º

C

-10

º

C to 60

º

C

Slow

Fast

Slow

Fast

300 mV

20 to 50 Hz

3%+.25 mV

6%+.5 mV

3.5%+.25 mV

7%+.5 mV

50 to 150 Hz

0.4%+.25 mV

1%+.5 mV

0.5%+.25 mV

1.5%+.5 mV

150 Hz to 10 kHz

0.3%+.25 mV

1%+.5 mV

0.4%+.25 mV

1.5%+.5 mV

10 kHz to 20 kHz

0.4%+.25 mV

1%+.5 mV

0.7%+.25 mV

1.5%+.5 mV

20 kHz to 50 kHz

2%+.3 mV

3%+.5 mV

3%+.3 mV

4%+.5 mV

50 kHz to 100 kHz

5%+.5 mV

5%+1 mV

7%+.5 mV

8%+1 mV

3V

20 to 50 Hz

3%+2.5 mV

6%+5 mV

3.5%+2.5 mV

7%+5 mV

50 to 150 Hz

0.4%+2.5 mV

1%+5 mV

0.5%+2.5 mV

1.2%+5 mV

150 Hz to 10 kHz

0.3%+2.5 mV

1%+5 mV

0.4%+2.5 mV

1.2%+5 mV

10 kHz to 20 kHz

0.4%+2.5 mV

1%+5 mV

0.5%+2.5 mV

1.2%+5 mV

20 kHz to 50 kHz

1%+3 mV

1.5%+6 mV

1.5%+3 mV

2%+6 mV

50 kHz to 100 kHz

2%+5 mV

3%+10 mV

3%+5 mV

4%+10 mV

30V

20 to 50 Hz

3%+25 mV

6%+50 mV

3.5%+25 mV

7%+50 mV

50 to 150 Hz

0.4%+25 mV

1%+50 mV

0.5%+25 mV

1.2%+40 mV

150 Hz to 10 kHz

0.3%+25 mV

1%+50 mV

0.5%+25 mV

1.2%+40 mV

10 kHz to 20 kHz

0.4%+25 mV

1%+50 mV

0.5%+25 mV

1.2%+40 mV

20 kHz to 50 kHz

1%+30 mV

1.5%+60 mV

1%+30 mV

2%+50 mV

50 kHz to 100 kHz
V<20V

2%+50 mV

3%+100 mV

2.5%+50 mV

4%+100 mV

150/300V

20 to 50 Hz

3%+.25V

6%+.50V

3.5%+.25V

7%+.50V

50 to 150 Hz

0.4%+.25V

1%+.5V

0.5%+.25V

1.2%+.40V

150 Hz to 2 kHz
V*Hz>2*106

0.3%+.25V

1.2%+.5V

0.5%+.25V

1.4%+.40V

2 kHz to 20 kHz
V<100V

0.4%+.25V

1.6%+.5V

0.5%+.25V

1.8%+.40V

20 kHz to 50 kHz
V<40V

1%+.30V

2%+.6V

1.2%+.30V

2.5%+.50V

* Sine wave inputs>6% of scale. Accuracy for signals with crest factors <2.