Fluke NetDAQ 2645A User Manual
Page 179

Maintenance
Performance Test
6
6-7
Table 6-2. Recommended Test Equipment
Instrument Type
Minimum Specifications
Recommended Model
Multifunction
DC Voltage:
Fluke 5700A
Calibrator
Range: 90 mV to 300V dc
Accuracy: 0.002%
AC Voltage:
Frequency
Voltage
Accuracy
1 kHz
29 mV to 300V
0.05%
100 kHz
15 mV to 300V
0.5%
Frequency:
10 kHz
1V rms
0.01%
Ohms:
Ohms
Accuracy
190
Ω
0.005%
1.9 k
Ω
0.005%
19 k
Ω
0.005%
190 k
Ω
0.005%
1.9 M
Ω
0.005%
Mercury Thermometer
0.02
°
C Resolution
Princo ASTM-56C
Thermocouple Probe
Type T
Fluke P20T
Oil/Water Bath
Thermos bottle and cap
Digital Multimeter
General Purpose Measurement
Fluke 77
Signal Generator
Sine wave. 0.5 to 1V rms, 10 Hz to 5 kHz
Philips PM5193
Alternative Equipment List
Instrument Type
Recommended Model
DC Voltage Calibrator
Fluke 5440B
DMM Calibrator
Fluke 5100B (AC Volts Only)
Function/Signal Generator
Philips PM5139 or HP33120A
Decade Resistance Source
General Resistance RDS 66A