Specifications

2640A Specifications

A

A-13

2640A 2-Wire Resistance Measurement Specifications

A-16.

The 2640A specifications for the 2-wire resistance measurement function is based
on the 4-wire resistance measurement specification (above) except you add a
nominal 5-Ohm (10-Ohm maximum) positive offset. This value varies for each
channel and with temperature (nominal +1%/ºC).

2640A RTD’s 4-Wire, per ITS-1990 Measurement Specifications

A-17.

Tables A-18 and A-19 provide 2640A specifications for the 4-wire Resistance-
Temperature Detector (RTD) measurement function. The 4-wire measurements
use 2 input channels a decade apart, e.g., channels 4 and 14.

Table A-18. 2640A 4-Wire RTD Temperature Coefficient

Specification

Characteristic

Temperature Coefficient

To calculate RTD accuracy for temperatures
between 28

°

C and 60

°

C, or -10

°

C and 18

°

C,

use a linear interpolation between the two
applicable points. e.g. if the applicable spec at
28

°

C is .2 and the spec at 60

°

C is .75, then the

specification at 40

°

C = (.75-.2)x(40-28)/(60-

28)+.2=.55*(12/32) + .2 = .406.

Accuracy at -20

°

C

Multiply the -10

°

C to + 60

°

C accuracy

specification by 2. After 1 hour warm-up. For
accuracy between -10

°

C and -20

°

C, interpolate

linearly.

Table A-19. 2640A 4-Wire RTD Specifications

Accuracy, 3

σ

±

ºC

Temperature

Resolution

90 Day, 18

º

C to

28

º

C

1 Year,

18

º

C to

28

º

C

1 Year, -10

º

C to

60

º

C

Slow

Fast

Slow

Fast

Slow

Slow

Fast

-200

º

C (Note1)

0.003

º

C

0.007

º

C

0.06

º

C

0.16

º

C

0.09

º

C

0.33

º

C

0.63

º

C

0

º

C (Note 1)

0.003

º

C

0.007

º

C

0.09

º

C

0.20

º

C

0.13

º

C

0.53

º

C

0.86

º

C

100

º

C

0.003

º

C

0.007

º

C

0.10

º

C

0.23

º

C

0.16

º

C

0.63

º

C

0.97

º

C

300

º

C

0.003

º

C

0.007

º

C

0.14

º

C

0.3

º

C

0.21

º

C

0.83

º

C

1.2

º

C

600

º

C

0.003

º

C

0.007

º

C

0.19

º

C

0.53

º

C

0.30

º

C

1.2

º

C

1.6

º

C

1. The Custom 385 range is not recommended for use below -80

º

C. (See Appendix D)