2640a dc voltage measurement specifications – Fluke NetDAQ 2645A User Manual
Page 218

2640A/2645A NetDAQ
Users Manual
A-8
2640A DC Voltage Measurement Specifications
A-13.
Tables A-9 to A-11 provide 2640A specifications for the dc voltage measurement
function.
Table A-9. 2640A DC Voltage Measurement General Specifications
Specification
Characteristic
Input Impedance
100 M
Ω
in parallel with 150 pF maximum for
ranges
≤
3V
10 M
Ω
in parallel with 100 pF maximum for ranges
>3V
Normal Mode Rejection
50dB minimum at 50 Hz/60 Hz
±
0.1%, Slow Rate
Common Mode Rejection
120dB minimum at dc, 50 Hz/60 Hz
±
0.1%, 1 k
Ω
imbalance, Slow Rate
80dB minimum at dc, 50 Hz/60 Hz
±
0.1%, 1 k
Ω
imbalance, Medium and Fast Rates
Channel-to-Channel Crosstalk
120dB minimum Slow Rate (e.g., 30V dc on
channel 1 may cause a 30µV error on channel 2)
100dB minimum Medium and Fast Rates (e.g., 1V
dc on channel 1 may cause a 10 µV error on
channel 2)
Temperature Coefficient
Add 1/10th the 90-day specification per
°
C above
28
°
C or below 18
°
C. (Generally, only the %input
portion is affected.)
Accuracy at -20
°
C
Multiply the -10
°
C to + 60
°
C accuracy specification
by 2. After 1 hour warm-up. For accuracy between
-10
°
C and -20
°
C, interpolate linearly.
Maximum Input Voltage
150V (300V for channels 1 and 11) to any input
terminal.
Table A-10. 2640A DC Voltage Range and Resolution Specifications
Resolution
Range
Slow
Fast
90 mV
.3 µV
1 µV
300 mV
1 µV
3 µV
3V
10 µV
30 µV
30V
100 µV
300 µV
150V/300V
1 mV
3 mV
Note 300V range applies to channels 1 and 11 only.