Power Quality Logger

Technical

Specifications

53

Power (Logging Functions A, P only) P, S, |P|

Active power P

As per EN 61036, class 2

Distorting power D

As per EN 61268, class 2 (A-version only)

Max value

Highest value per interval

Min value

Smallest value per interval

Phase uncertainty

< 0.3 degrees

Conditions Conductor

centered

within clamp jaws or

Flexi Set

Harmonics

V

m

, I

m

, THDV, THDI per IEC/EN 61000-4-7, class B

Voltage harmonics (function A)
intrinsic uncertainty:

For V

m

< 3% V

N

: < 0.15% V

N

For V

m

≥ 3% V

N

: < 5% V

m

Current harmonics (function A)
intrinsic uncertainty

For I

m

< 10% I

N

: < 0.5% I

N

For I

m

≥ 10% I

N

: < 5% I

m

THDV (function A) intrinsic
uncertainty at V

N

For THDV < 3%: < 0.15%

For THDV ≥ 3%: < 5%

THDV (function P) intrinsic uncertainty
at V

N

For THDV < 3%: < 1%

For THDV ≥3%: < 5%

THDI (functions A, P) intrinsic
uncertainty at I

I

For THDI < 3%: < 2%

For THDI ≥ 3%: < 5%

Statistics

Frequency

42 classes for 10 s mean values

Ripple-control signals

Interharmonics

21 classes for 3 s mean values

Analysis of logging data

Programming and analysis is done by PQ Log software on PC.