Users Manual

34

THDV – In Function A

Function A:

1

50

2

2

V

V

V

THD

n

n

=

=

V

n

: RMS value of harmonic frequency #n.

V

1

: RMS value of the fundamental frequency.

THDV: total contents of harmonics of the line power voltage as a percentage
of the fundamental.

This algorithm is according to EN 61000-4-7.

THD of currents:

Function A:

1

50

2

2

I

I

THDI

n

n

=

=

and

( )

=

=

50

2

2

n

n

I

A

THDI

I

n

: RMS value of harmonic frequency #n.

I

1

: RMS value of the fundamental frequency.

THDI: total contents of harmonics of the current as a percentage of the
fundamental.