Real configuration parameters – Hardy HI 1756-nDF User Manual
Page 38

34
●
●
●
●
●
Chapter 4
Parameter 12
Flow Rate Period
Used to set when the flow rate calculation is active. When set to 0 (default), the flow rate
calculation continuously runs, and when set to 1 the flow rate calculation is run only
during a fill/dispense cycle
Parameter 13
steady period
This parameter sets the number of samples that are processed to provide the weight
output.
The valid array sizes are 0, 16, 32, 64, and 128 (default).
When this value is set to zero the steady weight algorithm is bypassed and the weight
data value is used directly.
Parameter 14
auto preact enable
Enables the auto-preact function
Parameter 15
capture averages
Sets the number of fill/dispense cycles to average the target error values and flow rate
values to provide the auto_preact value.
REAL Configuration Parameters
Parameter
Name
Parameter
Number
Parameter
Default
Description
Span Weight
16.0
4,000
The total amount of weight, in selected
weight units, placed on the scale for the
high point when performing a “Traditional
Calibration”.
Cal low weight
17.0
0.0
Amount of test weigh, in selected weight
units, used, normally zero, for the low
point during hard calibration or reference
point during C2 calibration.
Auto Zero
tolerance
18.0
4.0
If used, the gross weight measurement
value will be set to zero whenever the
gross weight is within the auto zero
tolerance and the scale is not in motion.
Motion
tolerance
19.0
2.0
Value, in selected weight units, used to
determine if the scale is in motion.
Zero tolerance
20.0
4.0
Sets the weight range, in selected weight
units, around zero that will be accepted as
zero by the instrument.
Tare weight
21.0
0.0
The weight value, in selected weight units,
removed by the TARE function.
Start refill
weight
22.0
0.0
Weight reference which will cause a refill
cycle to automatically start
Stop refill
weight
23.0
4,000.0
Weight at which a refill cycle will
automatically stop