ITC Hemochron Response Whole Blood Coagulation System User Manual
Page 3

3
The system (Figure 1) contains a dual-well patented clot detection module. User interface is through a
keypad and display panel. Test results are shown on the display panel on completion of the test, and can be
printed.
Figure 1. HEMOCHRON Response Whole Blood Coagulation System
Two RS232 serial ports and a Centronix parallel port are included so that results and other information
from the data storage module can be downloaded to the laboratory computer or printed elsewhere. The
RS232 serial ports can also be used to connect an external barcode reader for importing Patient ID (PID)
and/or Operator ID (OID).
Definitions and Terms
The following acronyms and abbreviations are used in this manual, instrument screens and printouts:
ACT
Activated Clotting Time
APTT
Activated Partial Thromboplastin Time
DB
Database
ESV
Electronic System Verification
HRDM
HEMOCHRON Response Data Manager software program
idms
Integrated Data Management System
INR
International Normalized Ratio
LQC
Liquid Quality Control
OID
Operator Identification Number
PIN
Operator Personal Identification Number
PID
Patient Identification Number
POCC
Point of Care Coordinator
PPID
Fibrinogen Product Performance ID Code
PT
Prothrombin Time
QC
Quality Control
TVT
Temperature Verification Tube
Overview of Operation
Important: Disposable ready-to-use test tubes for use with the system can be obtained from ITC. Test
tubes from other manufacturers can be run, but the test being run will not be identified and the
clotting time results may be significantly different. The laboratory should verify performance if test
tubes from other manufacturers are used.
The system rotates the test tube at a constant rate while continuously monitoring the contents. An
integrated test type barcode reader decodes the test name and expiration date imprinted on the tube label.