Commissioning, Degas) – IKA C 1 Package 1/10 User Manual
Page 7

7
Hoses
• Discharge hose
• Inflow conduit
• Return conduit
• Venting hose
• Connection tube
SW8
Cooler
C1
C1
Cooler
C1
C1
(O2)
C1
(degas)
(water in)
(water out)
(overflow)
Filtersieb 10μm
Cooler
Exhaust
air
Pressure reducer C29
Adapter
SW10
Commissioning
Place of installation
A constant ambient temperature is an important requirement
for ensuring the high measuring accuracy of the system. The
following conditions must therefore be fulfi lled at the place of
installation:
•
No direct solar radiation
•
No draughts (e.g. beside windows, doors, air conditioning)
•
Suffi cient distance to radiators and other heat sources
•
The minimum distance between the wall and the rear side
of the device must be at least 25 cm.
•
The system must not have laboratory hardware such as
shelves, wire ducts, ring lines, etc. installed above it.
•
The (constant) room temperature should be around
20 °C ... 25 °C.
•
The system must be set up on a horizontal surface.
For operation of the system the following must be available at the
place of installation:
- A power supply corresponding to the nameplates on the system
components,
- An oxygen supply (99.95 % pure oxygen, quality 3.5; pressure
30 bar) with pressure gauge.
A shut-off valve for the oxygen supply must be installed. Observe
the instructions on handling oxygen given in the chapter „Safety
instructions“.
Observe the instructions on handling oxygen given in the chapter
„Safety instructions“. Please read these operating instructions
carefully. IKA
®
consider themselves responsible for the safety,
reliability and performance of the appliance only when
• the device has been operated in accordance with the
operating instructions
• the specifi cations for the place of installation have been met.
max. 1,5 bar
C1 092014