Default settings, Settings date / time, Units – IKA C 1 Package 1/10 User Manual
Page 30

30
•
C-value IB1; C-value IB2:
C value for the corresponding inner vessel
•
ID IB1; ID IB2:
Identification number of the inner vessel
•
Ref Energy:
Calorific reference value of benzoic acid
•
H
O
Combustible crucible:
Calorific value of the combustible crucible
•
Start temperature:
Changing the mode of operation between 22 ° C and 30 ° C.
•
QExtern 1; QExtern 2:
Default values for the external energies 1 and 2
•
O
2
rinse:
The inner vessel can before the measurement of oxygen
be rinsed (2x with subsequent filling)
•
Cool down:
Cooling the system after measurement
•
Extend. O
2
Emptying [min]:
0 - 99 adjustable
Settings
Date / Time
The setup menu contains the following menu items
•
Date / Time
•
Units
•
Display / Keypad
•
Language
•
Accessories
•
Chiller temperature
Setting the system time and date.
This data is used for automatic generation of a new measurement
name.
Units
The user can choose between the listed below unity
representations:
•
J/g
•
cal/g
•
BTU/lb
•
MJ/kg
Default settings
NOTE:
When using a gas-washing station, the ventilation time
must be extended.