IKA C 1 Package 1/10 User Manual
Page 23

23
•
Phase 9:
The system is completely emptied, and can be opened.
The inner vessel can be taken down. Through a visual check
is necessary to ensure that the sample in the crucible was
combusted completely.
After each combustion are all parts of the inner vessel that come
in contact with the combustion products to clean condensate.
It is sufficient to parts with an absorbent, lint-free cloth to wipe
off. Combustion residues in the inner vessel, for example, soot
or ash are eliminated in the same way. Can not be cleaned in
the manner described (eg due to burn-in or pitting corrosion) of
Cleaning the inner vessel
the inner vessel, it shall be subjected to no cases of mechanical
cleaning.
Please contact in such cases, contact your local service center or
send the inner vessel for inspection and cleaning to our factory.
Follow-up measurement
By pressing the multi-function element, the result display returns
to the measurement sub-menu and the next attempt can be
started.
Use of a disposable crucible
The disposable crucible can optionally be used instead of metal
or quartz crucible and burned during the measurement complete
(see chap. „Accessories“).
INFO: 50Joule omitted. With the combustible crucible not a
cotton thread is used.
Display the storage overview
If after pressing the „New Measurement“ button, the memory
overview is displayed, the capacity of the storage spaces for
measurements / calibrations are almost depleted. If the maximum
number is reached, no new measurements / calibrations more can
be saved for this mode. By pressing the OK key to access the screen
„New Measurement“. Delete measurements / calibrations are no
longer needed.
Result Display
NOTE:
Because the calorimeter C1 (all internal parts) still not in a
first measurement at operating temperature, the result of
the first measurement can differ significantly from the
subsequent measurements of a test series.
We therefore recommend not to evaluate the outcome and
possibly to delete.
Alternatively, a system test can be performed,
see chap. „System Test“.
C1 092014