Pressure test oxygen pressure, System check – IKA C 1 Package 1/10 User Manual
Page 13

13
Pressure test
Oxygen pressure
Information screen: oxygen pressure is too low.
NOTE:
The calorimeter C1 required for functionality an oxygen
pressure of at least 20 bar.
Reproducible measurements for a working pressure
of 30 bar is required.
System check
System start when the device is closed.
The system is brought through the system check in a safe
and completely deflated state. For this, the oxygen and water
discharge is performed.
Here, the following information screen is displayed:
INFO: During the system check detected no oxygen or the oxygen
pressure is too low, the following error message is displayed:
A pressure test must be performed after a decomposition vessel
has reached or exceeded the recommended number of ignition
cycles.
Once the pressure test has been performed a release code can be
entered to enable the unit to be used for further measurements.
The warning message will then disappear. The enable code can
be in the following screen, by pressing the „Next button“ to be
entered.
INFO: Using the unit may be further worked.
Contact for performing the pressure test, the
service department from IKA Works.
Observe the safety instructions!
C1 092014