HT instruments SPEED418 User Manual
Page 69
400
Series
EN - 68
12.2. SAFETY SPECIFICATION
12.2.1. General
Instrument safety:
IEC/EN61010-1, IEC / EN61557-1, -2, -3, -4, -6, -7
Technical documentations:
IEC/EN61187
Accessory safety:
IEC/EN61010-031, IEC/EN61010-2-032
Insulation:
double insulation
Pollution level:
2
Max height of use:
2000m (6562ft)
Overvoltage category:
CAT III 240V to earth, max 415V among inputs P, N, PE
5V to ground, max 7.2V
peak to peak
between pins of In1 Input
12.2.2. Reference standards for verification measurements
LOW
(200mA):
IEC/EN61557-4
M
: IEC/EN61557-2
RCD: IEC/EN61557-6
LOOP P-P, P-N, P-PE:
IEC/EN61557-3
Ra 15
mA
IEC/EN61557-3
123: IEC61557-7
12.2.3. AUX
Sound level measurement:
IEC/EN60651:1994/A1 type 1, IEC/EN60804:1994/A2 type 1
12.3. GENERAL CHARACTERISTICS
Mechanical data
Dimensions (L x W x H):
235 x 165 x 75mm (9 x 6 x 3in)
Weight (batteries included): 1250g (44ounces)
Power supply
Battery type:
6 batteries 1.5 V – LR6 – AA – AM3 – MN 1500
Low battery indication:
the symbol
is displayed when the battery voltage
is too low
Battery life:
approx >600 tests in all safety tests
approx 48 hours in PWR function
Auto power off:
if set to ON after about 5 minutes from the last key press or
measurement the instrument turns off automatically
Miscellaneous
Display:
LCD custom with backlight 73x65 mm
Memory:
500 memory locations
PC interface:
optical port
12.4. ENVIRONMENT
12.4.1. Environmental working conditions
Reference temperature:
23° ± 5°C (73°F ± 41°F)
Working temperature:
0 ÷ 40°C (32°F ÷ 104°F)
Relative humidity allowed:
<80%HR
Storage temperature:
-10 ÷ 60°C (14°F ÷ 140°F)
Storage humidity:
<80%HR
This instrument satisfies the requirements of Low Voltage Directive 2006/95/EC
(LVD) and of EMC Directive 2004/108/EC
12.5. ACCESSORIES
See packing list.