4. Once calibration is comple-
ted, in case the detected 
value is lower than 5
, the
instrument gives a double 
acoustic signal which signals 
the positive result of the test 
and displays a screen similar 
to the one reported here to 
the side
L O W
- - - -
R +
R -
- - - -
 - - - - 
- - - m A
- - - m A
CAL
4.00
0.21
Value of the calibrated 
resistance 
Func
Lim CAL
5. In order to delete the compensation resistance value of the cables, it is necessary to
perform a cable calibration procedure with a resistance higher than 5
 at test leads
(e.g. with open test leads).
 
6.2.2. Description of anomalous results 
1. By using the AUTO, R+ or R-
mode, in case the detected 
value is higher than the set 
limit, the instrument gives a 
long acoustic signal and 
displays a screen similar to 
the one reported here to the 
side
L O W
5 . 9 2
R +
R -
5 . 9 2
 5 . 9 1 
2 1 0 m A
2 1 0 m A
R > LIM
AUTO
4.00
0.21
Func
Lim CAL
2. By using the AUTO, R+ or R-
mode, in case the detected 
value is higher than the full 
scale, the instrument gives a 
long acoustic signal and 
displays a screen similar to 
the one reported here to the 
side
L O W
> 9 9 . 9
R +
R -
> 9 9 . 9
 > 9 9 . 9 
- - - m A
- - - m A
I < 200mA
AUTO
4.00
0.21
Func
Lim CAL
3. By using the AUTO, R+ or R-
mode, if the instrument 
detects a resistance which 
prevents a current of 200mA 
to circulate, it gives a long 
acoustic signal and displays 
a screen similar to the one 
reported here to the side
L O W
2 0 . 0
R +
R -
2 0 . 0
 2 0 . 0 
1 5 7 m A
1 5 7 m A
I < 200mA
AUTO
4.00
0.21
Func
Lim CAL
4.
The results displayed can be saved by pressing the SAVE key twice or the 
SAVE
key and, subsequently, the ENTER key (§ 9.1)