beautypg.com

Accu-Chek insulin pump User Manual

Page 47

background image

Chapter 2: Getting Started

ACCU-CHEK Spirit Insulin Pump

46

Remove the plunger rod by turning it counter-clockwise
out of the plunger.

Remove the filling needle with its protective cap from the
cartridge by turning counter-clockwise. Dispose of properly.

Push the cartridge protective cap tightly onto the cartridge
tip until it clicks.

The cartridge is now ready to use.

!

Air bubbles in the cartridge and infusion set may cause
the infusion of air instead of insulin. In this case, your body
does not receive the required amount of insulin and an
error E4: OCCLUSION may be displayed. Remove bubbles
while filling the cartridge and priming the infusion set and
without having the infusion set connected to your body.

Cold insulin may release air when warmed. Use only insulin
at room temperature when filling the cartridge and priming
the infusion set. Inspect the cartridge and the infusion set for
air bubbles at least every three hours during the day and
before you go to sleep. Remove any air bubbles and change
system components if needed.

Note:

Do not pull or push the plunger rod during removal.

CAUTION